Chapter 2 Command Systems
RIGOL
Programming Guide for DSA1000A series
2-97
Meas Setup, Avg Mode, Exp/Repeat
92.
[:SENSe]:TPOWer:MODE
Syntax
[:SENSe]:TPOWer:MODE AVERage|PEAK|RMS
[:SENSe]:TPOWer:MODE?
Function
Sets the type of the Time-domain Power measurement.
The query returns “AVER”, “PEAK” or “RMS”.
Default
PEAK
Front Panel
Meas, T-Power
Meas Setup, Peak/Average/RMS
93.
[:SENSe]:TPOWer:LLIMit
Syntax
[:SENSe]:TPOWer:LLIMit <time>
[:SENSe]:TPOWer:LLIMit?
Function
Sets the left limit of the Time-domain Power measurement, the default
unit is s, such as
:TPOW:LLIM 0.005
The query returns “5.000000E-03”.
Explanations
<time> ranges from 0 to Right limit.
Default
0 s
Front Panel
Meas, T-Power
Meas Setup, Start Line
94.
[:SENSe]:TPOWer:RLIMit
Syntax
[:SENSe]:TPOWer:RLIMit <time>
[:SENSe]:TPOWer:RLIMit?
Function
Sets the right limit of the Time-domain Power measurement, the
default unit is s, such as
:TPOW:RLIM 0.01
The query returns “1.000000E-02”.
Explanations
<time> ranges from the left limit to Sweep Time.
Default
100 ms
Front Panel
Meas, T-Power
Meas Setup, Stop Line
Содержание DSA1000A Series
Страница 1: ...RIGOL Programming Guide DSA1000A Series Spectrum Analyzer May 2011 RIGOL Technologies Inc...
Страница 2: ......
Страница 14: ......
Страница 137: ...Chapter 3 Programming Examples RIGOL Programming Guide for DSA1000A series 3 3...
Страница 154: ...RIGOL Chapter 3 Programming Examples Programming Guide for DSA1000A series 3 20 6 Exit code...
Страница 155: ...Chapter 3 Programming Examples RIGOL Programming Guide for DSA1000A series 3 21 7 Execute results...
Страница 158: ...RIGOL Chapter 3 Programming Examples Programming Guide for DSA1000A series 3 24...