Chapter 2 Command Systems
RIGOL
Programming Guide for DSA1000A series
2-89
Syntax
[:SENSe]:EBWidth:AVERage:TCONtrol EXPonential|REPeat
[:SENSe]:EBWidth:AVERage:TCONtrol?
Function
Sets the average mode of the Emission Bandwidth measurement.
The query returns “EXP” or “REP”.
Explanations
In Exp mode, the result is from the exponent average calculation
of the current N (specified by
[:SENSe]:EBWidth:AVERage:COUNt) measurements.
In REP mode, the result is the arithmetic average of the current N
(specified by [:SENSe]:EBWidth:AVERage:COUNt) measurements.
Default
EXPonential
Front Panel
Meas, EBW
Meas Setup, Avg Mode, Exp/Repeat
64.
[:SENSe]:EBWidth:MAXHold:STATe
Syntax
[:SENSe]:EBWidth:MAXHold:STATe OFF|ON|0|1
[:SENSe]:EBWidth:MAXHold:STATe?
Function
Turns on or off the Max Hold function.
The query returns “0” or “1”.
Default
OFF|0
Front Panel
Meas, EBW
Meas Setup, Max Hold, On/Off
65.
[:SENSe]:EBWidth:FREQuency:SPAN
Syntax
[:SENSe]:EBWidth:FREQuency:SPAN <freq>
[:SENSe]:EBWidth:FREQuency:SPAN?
Function
Sets the span of the Emission Bandwidth measurement, the default
unit is Hz, such as
:EBW:FREQ:SPAN 1000000 or :EBW:FREQ:SPAN 1MHZ
The query returns “1000000”.
Explanations
<freq> ranges from 100 Hz to 3 GHz.
Default
2 MHz
Front Panel
Meas, EBW
Meas Setup, Span
66.
[:SENSe]:EBWidth:XDB
Syntax
[:SENSe]:EBWidth:XDB <real_amp>
[:SENSe]:EBWidth:XDB?
Содержание DSA1000A Series
Страница 1: ...RIGOL Programming Guide DSA1000A Series Spectrum Analyzer May 2011 RIGOL Technologies Inc...
Страница 2: ......
Страница 14: ......
Страница 137: ...Chapter 3 Programming Examples RIGOL Programming Guide for DSA1000A series 3 3...
Страница 154: ...RIGOL Chapter 3 Programming Examples Programming Guide for DSA1000A series 3 20 6 Exit code...
Страница 155: ...Chapter 3 Programming Examples RIGOL Programming Guide for DSA1000A series 3 21 7 Execute results...
Страница 158: ...RIGOL Chapter 3 Programming Examples Programming Guide for DSA1000A series 3 24...