50 Ω Input Impedance
Channel
Vertical Scale
Test Result
Limit
Pass/Fail
CH1
100 mV/div
48.75 Ω to 51.25 Ω
500 mV/div
48.75 Ω to 51.25 Ω
CH2
100 mV/div
48.75 Ω to 51.25 Ω
500 mV/div
48.75 Ω to 51.25 Ω
CH3
100 mV/div
48.75 Ω to 51.25 Ω
500 mV/div
48.75 Ω to 51.25 Ω
CH4
100 mV/div
48.75 Ω to 51.25 Ω
500 mV/div
48.75 Ω to 51.25 Ω
5.2
DC Gain Accuracy Test
5.2.1
Specification
DC Gain Accuracy
Specification
±2% of Full Scale
[1]
NOTE
[1]: Full scale = 8 × Current Vertical Scale. 1 mV/div and 2 mV/div are a magnification of 4
mV/div setting. For vertical accuracy calculations, use full scale of 32 mV for 1 mV/div and 2
mV/div vertical sensitivity setting.
5.2.2
Test Connection Diagram
Fluke 9500B
CH1
DS70000
5.2.3
Test Procedures
WARNING
Before connecting, disconnecting, or moving the test hookup, disable the output of the
signal generator to avoid causing the dangerous voltage.
Performance Verification Test
DS70000 Performance Verification
14