3. Turn on Fluke 9500B; set its impedance to 1 MΩ and select the resistance
measurement function. Read and record the resistance measured.
4. Adjust the vertical scale of CH1 of the oscilloscope to 500 mV/div; read and record
the resistance measured.
5. Turn off CH1. Measure the resistances of CH2, CH3, and CH4 respectively using the
method above and record the measurement results.
Impedance test when the input impedance is 50 Ω
1. Connect the active head of Fluke 9500B to CH1 of the oscilloscope, as shown in the
figure above.
2. Configure the oscilloscope:
a. Use the method illustrated above to turn on CH1.
b. Set the input impedance of CH1 to 50 Ω.
c. Set the vertical scale of CH1 to 100 mV/div.
3. Turn on Fluke 9500B; set its impedance to 50 Ω and select the resistance
measurement function. Read and record the resistance measured.
4. Adjust the vertical scale of CH1 of the oscilloscope to 500 mV/div; read and record
the resistance measured.
5. Turn off CH1. Measure the resistances of CH2, CH3, and CH4 respectively using the
method above and record the measurement results.
5.1.4
Test Record Form
1 MΩ Input Impedance
Channel
Vertical Scale
Test Result
Limit
Pass/Fail
CH1
100 mV/div
0.99 MΩ to 1.01 MΩ
500 mV/div
0.99 MΩ to 1.01 MΩ
CH2
100 mV/div
0.99 MΩ to 1.01 MΩ
500 mV/div
0.99 MΩ to 1.01 MΩ
CH3
100 mV/div
0.99 MΩ to 1.01 MΩ
500 mV/div
0.99 MΩ to 1.01 MΩ
CH4
100 mV/div
0.99 MΩ to 1.01 MΩ
500 mV/div
0.99 MΩ to 1.01 MΩ
Performance Verification Test
Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved.
DS70000 Performance Verification
13