5
Performance Verification Test
This chapter takes DS70504 as an example to illustrate the performance verification
test methods and procedures of DS70000 series digital oscilloscope. The
recommended test devices in this manual is shown below. You can also use other
devices that fulfill the "Specification" in the following table.
Table 5.1 Test Devices
Device
Specification
Recommended Model
DC Votage
Source
3 mV to 4 V, ±0.1% accuracy
Fluke 9500B
RF Signal
Generator
10 kHz to 5 GHz
RIGOL DSG3000B Series RF
Signal Generator (with the
OCXO-B08 option)
Power Meter
1 MHz to 5 GHz, ±3% accuracy
-
Power Sensor
1 MHz to 5 GHz, ±3% accuracy
-
Power Splitter
Full-scale accuracy ≤ 0.2 dB
-
TIP
1. Make sure that the oscilloscope passes the self-test and self-calibration is performed before
executing the performance verification tests.
2. Make sure that the oscilloscope has been warmed up for at least 30 minutes before
executing any of the following tests.
3. Please reset the instrument to the factory setting before or after executing any of the
following tests.
5.1
Impedance Test
5.1.1
Specification
Input Impedance
Analog Channel
1 MΩ: 0.99 MΩ to 1.01 MΩ
50 Ω: 48.75 Ω to 51.25 Ω
Performance Verification Test
Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved.
DS70000 Performance Verification
11