RIGOL
Chapter 6 : Reference
Appendix A: Specifications
All specifications apply to the DS5000- Series Digital Oscilloscopes and a probe with
the Attenuation switch set to 10X unless noted otherwise. To meet these
specifications, two conditions must first be met:
The instrument must have been operating continuously for twenty minutes
within the specified operating temperature.
You must perform the Self Cal operation, accessible through the Utility menu, if
the operating temperature changes by more than 5 ° C.
All specifications are guaranteed unless noted “typical”.
Specifications
Acquisition
Sampling Modes
Real-Time
Equivalent
Sampling Rate
1GSa/s (DS5000CA/DS5000MA)
250MSa/s
(
DS5000C/DS5000M
)
50GS/s
Averages
N time acquisitions, all channels simultaneously, N is
selectable from 2, 4, 8, 16, 32, 64, 128 and 256
Probe
1X position
10X position
Bandwidth
DC to 6 MHz
DC to full bandwidth
Attenuation ratio
1:1 10:1
Compensation
Range
10pf-35pf Oscilloscope with 1M
Ω
input
Input resistance
1M
Ω
±2% 10M
Ω
±2%
Input capacitance
85pf-115pf 14.5pf-17.5pf
Input voltage
150V RMS CAT I or
150V DC CAT I
150V RMS CAT II or
150V DC CAT II
100V RMS CAT III or
100V DC CAT III
300V RMS CAT I or 300V DC
CAT I
300V RMS CAT II or 300V DC
CAT II
100V RMS CAT III or 100V DC
CAT III
Copyright 2004 Rigol Electronic
6-1
User Manual for DS5000 Series