DP700 Calibration Guide 

Document Overview 

 

This manual gives a detailed description of the calibration notices and calibration 

process of DP700 series programmable linear DC power supply (hereinafter referred 

to as "DP700 series"). 

 

Main Topics in this Manual: 

 

Chapter 1 Calibration Notices 

This chapter gives a detailed description of the recommended calibration interval, 

calibration safety, how to stop the calibration, how to acquire the calibration service, 

the recommended calibration devices, and the precautions during calibration. 

 

Chapter 2 Calibration Process 

This chapter gives a detailed description of the calibration page of DP series and the 

calibration process for each calibration item. 

 

Format Conventions in this Manual: 

 

 

The function key on the front panel is denoted by the format of "Key Name (Bold) 
+ Text Box" in the manual. For example, On/Off denotes the "On/Off" key. 
 

 

Use the screen shot to indicate the key. For example, 

  denotes the Power 

key. 

 

Content Conventions in this Manual: 

 

DP700 series programmable linear DC power supply includes the following two 

models. Unless otherwise specified, this manual takes DP711 as an example to 

illustrate the calibration notices and calibration process of DP700 series. 

 

Model 

Number of Channels 

Output Voltage/Current 

DP711 

30 V/5 A 

DP712 

50 V/3 A 

Содержание DP700 Series

Страница 1: ...RIGOL Calibration Guide DP700 Series Programmable Linear DC Power Supply July 2016 RIGOL TECHNOLOGIES INC...

Страница 2: ......

Страница 3: ...on is subject to change without notice RIGOL shall not be liable for either incidental or consequential losses in connection with the furnishing use or performance of this manual as well as any inform...

Страница 4: ...alibration Process This chapter gives a detailed description of the calibration page of DP series and the calibration process for each calibration item Format Conventions in this Manual The function k...

Страница 5: ...I Chapter 1 Calibration Notices 1 1 Calibration Interval 1 1 Calibration Safety 1 1 To Stop the Calibration 1 2 To Acquire the Calibration Service 1 2 Recommended Calibration Devices 1 2 Calibration P...

Страница 6: ......

Страница 7: ...en configured with a default calibration password 123456 The password is stored in the internal non volatile memory and will not be changed at power off or remote interface reset You can also modify t...

Страница 8: ...ain in the next calibration CAUTION If you stop calibration when the instrument is writing the new calibration factors into the internal non volatile memory you may lose all the calibration factors an...

Страница 9: ...rform all the calibration procedures Ensure that the ambient temperature stays between 18 C and 28 C Ideally the calibration should be performed at 23 C 1 C Ensure that the ambient relative humidity i...

Страница 10: ......

Страница 11: ...a The detailed operation method for each step is as follows Step 1 Enter the calibration page 1 Warm up DP700 series for at least 30 minutes 2 Press System to enter the system utility function interfa...

Страница 12: ...ibration items The calibration items for DP700 series are available in Table 2 1 You can select one or multiple items for calibration RIGOL recommends that you perform the calibration for all the avai...

Страница 13: ...are reference voltage current values At this time use the num key or the knob for a negative value input use the knob to enter the measurement value of the digital multimeter For setting methods refer...

Страница 14: ...the currently calibrated data will not be saved and then you exit the calibration page Figure 2 4 Save the calibration data Dialog Box Note If you want to calibrate multiple items RIGOL suggests you c...

Страница 15: ...tive terminal of the digital multimeter with the terminal of the channel output of DP711 and connect the negative terminal of the digital multimeter with the terminal of the channel output of DP711 2...

Страница 16: ...ding to the ID sequence then repeat Step 5 and 6 until you complete the calibration operation for all the calibration points 8 Press the Down arrow key or use the knob to switch the parameter focus to...

Страница 17: ...to Auto 4 Enter the DP711 calibration page and select V ADC then press the Right arrow key The cursor switches to ID of the first calibration point 5 Press OK The Meas value measurement value of the...

Страница 18: ...inal of the digital multimeter with the terminal of the channel output of DP711 and connect the negative terminal of the digital multimeter with the terminal of the channel output of DP711 2 Press On...

Страница 19: ...ding to the ID sequence then repeat Step 5 and 6 until you complete the calibration operation for all the calibration points 8 Press the Down arrow key or use the knob to switch the parameter focus to...

Страница 20: ...and the scale to A 4 Enter the DP711 calibration page and select I ADC then press the Right arrow key The cursor switches to ID of the first calibration point 5 Press OK The Meas value measurement val...

Страница 21: ...ion point there is only one calibration point for OVP calibration 3 Press OK and then the On Off key on the front panel of DP711 is illuminated automatically Then the instrument starts to perform auto...

Отзывы: