RIGOL
Chapter 2 Front Panel Operations
2-4
DG800 User's Guide
To Set Frequency/Period
Frequency is one of the most important parameters of basic waveforms. For different
instrument models and waveforms, the setting ranges of frequency are different. The
default frequency is 1 kHz.
Sine
Square Ramp
Pulse
Noise (-3 dB)
DG812/DG811 1 μHz to
10 MHz
1 μHz to
5 MHz
1 μHz to
200 kHz
1 μHz to
5 MHz
100 MHz
Bandwidth
DG822/DG821 1 μHz to
25 MHz
1 μHz to
10 MHz
1 μHz to
500 kHz
1 μHz to
10 MHz
100 MHz
Bandwidth
DG832/DG831 1 μHz to
35 MHz
1 μHz to
10 MHz
1 μHz to
1 MHz
1 μHz to
10 MHz
100 MHz
Bandwidth
The frequency displayed on the screen is the default value or the frequency
previously set. When the instrument function is changed, if this frequency is valid
under the new function, the instrument will still use this frequency; otherwise, the
instrument would display a prompt message and set the frequency to the frequency
upper limit of the new function automatically.
Tap the Freq parameter input field to input the desired frequency value with the
numeric keypad, then select the unit. Tap "Ok".
The frequency units available are MHz, kHz, Hz, mHz, and μHz.
Tap the label again to switch to Period setting.
The period units available are s, ms, μs, and ns.
You can also use the arrow keys and the knob to set the parameter value: press
down the knob to enter the editing mode, use the arrow keys to move the cursor to
select the digit to be edited, and then rotate the knob to modify the value. Besides,
you can use the knob to switch between different parameters.
Содержание DG800 Series
Страница 2: ......
Страница 14: ......
Страница 30: ......
Страница 46: ...RIGOL Chapter 2 Front Panel Operations 2 16 DG800 User s Guide Figure 2 3 Sine Waveform...
Страница 57: ...Chapter 2 Front Panel Operations RIGOL DG800 User s Guide 2 27 Figure 2 5 Harmonic...
Страница 136: ......
Страница 142: ......
Страница 148: ...RIGOL Index 2 DG800 User s Guide VISA Address 2 94 Waveform Combination 2 89 Width Deviation 2 55...