Phoenix HT User Manual
© 2007–2012 Rigaku Automation, Inc.
12
401718 Rev A
2.2. Shutting Down the System
The power-down sequence for the
CrystalMation
Phoenix HT instrument is:
1. Click the
Disconnect
icon in the
Phoenix
software.
When asked if you wish to disconnect from the Phoenix RE software, click OK.
When asked if you wish to disconnect from the Art Robbins Phoenix software, click OK.
2. Push the red
Emergency Motor Off
(EMO) button.
3. Power off the Nano-Dispense Controller.
4. Power off the Phoenix HT
instrument.
5. Shut down the laptop computer via Windows.
6. Rock the power switch on the outside of the enclosure to the 0 position.
2.3. Running the System
Follow the operating procedures current in your laboratory setting. Refer to the guidelines and
instructions in the original-equipment
Phoenix
Setup Guide,
in “Chapter 3. Quick Start Procedures”
on page 14, and in the “Chapter 4. Phoenix HT Application Reference” on page 19 of this manual.
2.4. Status Light Indicators
The status lights for the Phoenix HT enclosure indicate the following conditions.
Color
Standard Definition
Machine Specific
All Off
System has no power
Solid Red
Power is on, software launched, software in
stopped state, or machine is not ready to process
orders. Machine can not start processing orders.
Stopped, not ready for plate to be
made.
Blinking Red
Power is on, software launched, system has an
error condition and system has stopped, or the
system requires initialization, or the EMO is
active.
Not initialized, EMO pressed, liquid
handling error.
Solid Amber
Power is on, software launched, no errors but
supplies are low and system will run out and stop
soon, or the temperature is out of specification.
Supplies low.
Blinking Amber
Power is on, software launched, no errors but
system has run out of supplies and stopped, or
the software is paused, or the doors are open or
unlocked.
Software paused, door open, low
water pressure.
Solid Green
Power is on, software launched, no errors,
software is in idle state, waiting on orders. The
machine will automatically process orders when
they become available.
Idle, waiting on orders.
Blinking Green
Power is on, software launched, no errors, system
is running, actively processing/moving.
Order being processed.
Blinking
Amber+Red
Power is on, software is not launched or control
system is not responding.
Application software not
connected.