RF602 [Revision 1.2.0] 28.01.2019
Laser Triangulation Sensors. RF602 Series
21
12.2. Voltage output 0…10 V
The connection scheme is shown in the figure. To reduce noise, it is recommended
to install RC filter before the measuring instrument. The filter capacitor value is indicated for
maximum sampling frequency of the sensor (9,4 kHz) and this value increases in proportion
to the frequency reduction.
12.3. Configuration parameters
12.3.1. Range of the analog output
While working with the analog output, resolution can be increased by using the
‘Window in the operating range’ function which makes it possible to select a window of
required size and position in the operating range of the sensor within which the whole range
of analog output signal will be scaled.
Note
. If the beginning of the range of the analog signal is set at a higher value than
the end value of the range, this will change the direction of rise of the analog signal.
12.3.2. Analog output operation mode
When using the ‘Window in the operating range’ function, this mode defines the
analog output operation mode.
Analog output can be:
·
in the window mode or
·
in the full mode.
"Window mode"
. The entire range of the analog output is scaled within the selected
window. Outside the window, the analog output is "0".
"Full mode"
. The entire range of the analog output is scaled within the selected
window (operating range). Outside the selected window, the whole range of the analog
output is automatically scaled onto the whole operating range of the sensor (sensitivity
range).
12.4. Factory parameters table
Parameter
Value
Range of the analog output
Measurement range of the sensor
Analog output operation mode
Window
13. Parameterization program
13.1. Function
The RF60Х-SP software is intended for:
1) Testing and demonstration of work of RF602 series sensors;
2) Setting of the sensor parameters;
3) Reception and gathering of the sensor data signals
Download link: