Tapes geometry measurement system  

 

  

 [Version 1.0]  08

th

  December 2014                                       

 

 

14 

Figure 16 

 

Figure 17 

When typing the corresponding values the keyboard appears automatically. 

11.  Intended use 

11.1.   Prepatation for use 

Preparation of the system includes: 

  external inspection; 

  set up on a conveyer; 

  switching the system; 

  calibrating the system; 

  system setup.

 

11.1.1. External inspection 

Before operating it is needed to ensure of the serviceability of the equipment, 

check the cables, ground wires. Check the condition of output windows of the scanner 
and, if necessary, wipe them with a soft cloth. 

 

11.1.2. Set up on a conveyer 

Install the system on a conveyer according to the Fig. 2-5 of this manual. Make 

the electrical connections in accordance with the Fig. 8. 

 

11.1.3. Switching the system 

11.1.4. Calibrating the system 

Выполнить калибровку системы в соответствии с п. 10.3.2 настоящего руководства. 
Калибровка установки производится: Calibrate the machine according to the Para-
graph 10.3.2. Calibration should be held: 

  on a weekly basis before the work shifts, 

 

  if the position of the system was changed, 

 

  when the ambient temperature has changed (several degrees in average 

per day in relation to the previous day).

 

11.1.5. Configuring system settings 

Configure the settings according to the Paragraph 10.4 and 10.5. 

11.1.6. Checking the system 

To check the functionality of the  system conduct a full cycle control of the geo-

metrical parameters of the control sample tape. It is recommended to conduct at list one 
functional test of the system in a month.

 

11.2.  Operating the system 

The geometric parameters measurement cycle is fully automated and operation 

of the system is reduced to the work with the program. 

  prepare the system according to the Paragraph 11.1. 

  to start the measurement process press the 

Start 

button

.

  

Содержание RF089 Series

Страница 1: ...AUTOMATED SYSTEM FOR TAPES GEOMETRY MEASUREMENT RF089 Series...

Страница 2: ...on of the thickness measurement micrometers 10 10 3 2 Thermometers calibration 11 10 4 Settings 11 10 5 Measurement journal 12 11 Intended use 14 11 1 Prepatation for use 14 11 1 1 External inspection...

Страница 3: ...ents EN 61326 1 2006 Electrical Equipment for Measurement Control and Labora tory Use EMC Requirements General requirements 3 Laser safety The micrometers make use LED or c w 660 nm wavelength semicon...

Страница 4: ...FROM SHAFT AND TAPE Figure 2 The tape thickness is calculated as the difference between the position of shad ow borders of the tape and the shaft The position of shadow border of the shaft is con trol...

Страница 5: ...er of temperature control points 3 Measurement range for the tape thickness mm 0 02 20 Accuracy of the thickness measurement um 5 Measurement range for the tape width mm 0 2 20 Accuracy of the width m...

Страница 6: ...ement system Version 1 0 08th December 2014 6 7 Overall and mounting dimensions 7 1 Measurement modules Fig 5 and 6 show the dimensions of measurement modules as well as their instal lation on equipme...

Страница 7: ...sults and store the data Fig 6 shows overall dimensions of the device Figure 6 8 Block diagram Fig 7 shows the block diagram of the system Optical micrometers RF651 are linked via RS485 interface Tem...

Страница 8: ...items The power should be switched off while mounting cable connections 10 Software 10 1 Main functions The software provides receipt and analysis of data from micrometers and temperature sensors with...

Страница 9: ...by the main micrometers see P 5 Shaft_cl position of the shaft surface main micrometers while system calibra tion see P 10 3 Shaft_sp the current position of the shaft surface indicated by subsidiary...

Страница 10: ...emove tapes from the shaft start measurement process by pressing Start open the calibration menu Settings Set etalon Sw State Fig 11 select the numbers of the micrometers to be calibrated numbers 7 15...

Страница 11: ...ph for two different temperatures By pressing the Set button the previous calibration value is over written Figure 12 10 4 Settings All sensors in the system are numbered Index from 1 to 18 where the...

Страница 12: ...journal To view the measurement log enter the Journal menu Fig 13 This menu is intended for viewing the measurements stored on a flash drive in the SQLite database file named keep db Each entry has t...

Страница 13: ...system Version 1 0 08th December 2014 13 Figure 13 Figure 14 NOTE To enter the date and time values are special graphics components shown in Figures 15 and 16 respectively To enter numeric values use...

Страница 14: ...3 Switching the system 11 1 4 Calibrating the system 10 3 2 Calibrate the machine according to the Para graph 10 3 2 Calibration should be held on a weekly basis before the work shifts if the positio...

Страница 15: ...cking of completeness checking for any damage of the structural elements power and instrument cables indicators and connectors weakening of screw connections and insulation failures before starting wo...

Отзывы: