5010e_513.doc / 22.01.10
10.4
SHORT SPECIFICATIONS
Type
Semi-automatic, single-beam filter photometer
Light Source
Halogen lamp – 12 V, 20 W with protection for lifetime
Wavelength
340 nm – 800 nm
Wavelength
Selection
Automatic via 9-position filter wheel :
6 standard interference filters: 340 nm, 405 nm, 492 nm, 546 nm, 578 nm
and 623 nm; 3 positions for optional filter of choice
Photometric
Range
0 – 2.5 A
Cuvette System
Micro flow cell:32
µ
l, 10 mm light path interchangeable with normal standard
cuvettes (macro or semi-micro, disposable or special optical glass)
Temperature Control
Internal Peltier element, temperature variable, pre-adjusted to 25 °C,
30 °C and 37 °C
Equilibration time for aspirated reaction mixture to reach 37 °C from
ambient temperature: 15 s
Aspiration System
Built-in peristaltic pump driven by stepper motor, programmable
aspiration volume controlled by infrared light barrier
Sipping Volume
Minimum
250
µ
l, typically 500
µ
l up to 2000
µ
l
Separate setting of aspirate volume and wash volume
Operator Interface
Touchscreen for direct functions and alphanumerical inputs
Data Presentation
Graphic display: White characters or symbols, blue background, lighted,
resolution 240 * 128 dots
Integrated Printer
Graphics printer, dot-matrix, 24 characters per line
Languages
English and German/Indonesian/Russian/Spanish
Memory
General operating software can be updated by PC
Reagent-open system with capacity for up to 231 programmable methods
Data import by touchscreen or PC
Up to 50 nonlinear calibration curves with max 20 sets of points can be
stored
Signal Port
Serial port for connection to an external printer and/or PC
Data Logging
Up to 2970 results of max 99 samples can be managed in the memory
Measurement
Procedures
Absorbance
End point with factor, standard or multiple standards,
with or without reagent blank and/or sample blank
Bichromatic end point
Kinetics with factor, standard or multiple standards, with or without blank
Fixed time with factor, standard or multiple standards,
with or without reagent blank
Turbidimetry with optional timer function
Single, double and triple determinations
Curve fitting for nonlinear standard curves
Free hemoglobin in combination with optional interference filters
Quality Control
Up to 50 methods can be controlled with two control serums,
Levey Jennings plot
Measuring Time
Kinetic: variable from 3 – 28 deltas, time per delta 4 – 255 s
Fixed time: variable from 0 – 65535 s
Delay Time
Programmable from 0 – 65535 s
Mains Supply
Range: 90 V
AC
up to 264 V
AC
at 50/60 Hz
Dimensions
Length 33 cm x width 34 cm x height 18 cm
Weight
6 kg
Marking
ROBERT RIELE GmbH & Co KG
Page 68
10 - TECHNICAL DATA