5010e_513.doc / 22.01.10
7.2.9
Temperature ON / OFF
7.2.9
Temperature ON / OFF
The current status of the tempering is indi-
cated in the first line by OFF or ON. Besides
the inserted cuvette adaptor is marked by D or
E.
Change setting by [SELECT]. Following
options are possible:
•
OFF
•
25°
C
•
30°
C
•
37°
C
Save setting permanently by [OK].
Save setting temporarily up to next switch-off
of the device by [EXIT].
T E M P E R A T U R E O N D
O F F
2 5 C
3 0 C
3 7 C
E X I T O K S E L E C T
7.2.10 Temperature
adjustment
The temperature control was
adjusted at the factory!
Anyhow the tempering control can be cali-
brated provided that the tempering was
switched on for at least 30 minutes:
Measure the current temperature with an
independent measuring system (e.g. thermis-
tor, REF 090-063) inside the cuvette and enter
this value. According to the difference to 25.0
°C, 30.0 °C or 37.0 °C the system corrects its
internal setting. The calibrating of the tem-
perature is interrupted when the tempering is
off or the temperature unstable.
Enter the password “5010”.
Enter the actual temperature in TEMP.
ACTUAL four-digit in °C (e.g. 36.90) and
confirm by [E].
T E M P . A D J U S T M E N T
T E M P . A C T U A L : 3 6 . 9 0
1 2 3 4 5 6 7 8
E S C 9 0 . + - E