8.2.4.2 REMOTE CONTROL
Activate remote control by [START].
When activated, Photometer
4040
+
can be
remote-controlled by a PC and a suitable
program.
Deactivate remote control by keeping the
touchscreen pressed for some time.
By [EXIT] the program returns to the
superordinate menu item.
8.2.4.3 SERIAL PRINTER ON / OFF
The current status of the external serial
printer is indicated in the first line by OFF or
ON.
Change setting by [SELECT]
Save setting permanently by [OK]
Save setting temporarily up to next switch-
off of the device by [EXIT]
8.2.4.4 BAUD RATE
The current baud rate is shown in the first
line.
Change setting by [SELECT].
Save setting permanently by [OK].
Discard changes by [EXIT].
8.2.6
Quality control
In Photometer
4040+
up to 50 methods can be supervised with a quality control. The device can manage up to 6
control serums. Each QC supervised method can be connected with 2 control serums. The QC data of a series
of measurements are stored in a daily memory. Each reading is stored with method number, date and user
identification. From the daily memory the individual QC data can become deleted or saved in the monthly memory
of the corresponding method. The monthly memory of a QC method can record up to 31 readings. With the 32nd
the oldest reading is deleted in the memory. For the calculation of the quality values of a method at least 20
readings in the monthly memory must be present. The average of all readings, the standard deviation and the
coefficient of variation are calculated. Contents of the daily and monthly memory can be indicated and printed out.
Except the basic methods all methods can be connected with a quality control. The method-typical data of a
control serum are entered via the method editor (see chapter 7 METHOD EDITOR).