[EDIT CURVE] After inquiry of the curve
number all curve bases can be edited (see
fig. 7.2.2.2).
[SHOW CURVE] After inquiry of the curve
number the function will be shown.
[LIST] The method number and the date of
preparation of all current curves are shown.
[EXIT] Return to the utilities
Fig. 7.2.2.2
[P+] and [P-] Consecutive numbering of the
current bases
[A/C] Switch input between A for
absorbance and C for concentration
[E] Accept the edited value
Input and confirmation of a single “0”
at A lead to the deletion of the current pair
of points. In order to set the value to zero
enter e.g. “0.0”.
[ESC] End input and save curve data. The
bases are sorted in ascending order
according to their A value.
8.2.3
Printer ON / OFF
The current status of the internal printer is
indicated in the first line by OFF or ON.
Change setting by [SELECT]
Save setting permanently by [OK]
Save setting temporarily up to next switch-
off of the device by [EXIT]
NO. ABS. RESULT
- - - - - - - - - - - -
1 0.675 19.8
TIME: 11:21:32
When printer ON via [MODE] [PRN]
after a measurement the current time can be
printed out by [TIME].
P R I N T E R O F F
E X I T O K S E L E C T
C U R V E N O . x x
P 1 A - 0 0 0 0 0 N E W A
C - 0 0 0 0 0
1 2 3 4 5 6 7 8
E S C 9 0 . P + P - A / C E
C U R V E
N O . 2 0
E X I T