4040+e_22.docx / 06.10.17
ROBERT RIELE GmbH & Co KG
Page 24
6.3
SURVEY OF THE METHODS
The calculation procedures, on which all methods are traceable from the list of methods, are mentioned in the
following table. Criterion is the characteristic of the calculation procedure (see below). For detailed description of
the respectively accompanying procedure of method see chapter 6.4 - DESCRIPTION OF METHOD
PROCEDURES.
CP-No.
Characteristic
Method
Calculation formula
CP 1
C/F
Endpoint with Factor
C = F
A
S
CP 2
C/F/Rb
Endpoint with Factor
C = F
(A
S
- A
RB
)
CP 3
C/F/Sb
Endpoint with Factor
C = F
|A
S
- A
SB
|
CP 4
C/F/SbRb
Endpoint with Factor
C = F
(|A
S
- A
SB
| - |A
RB
- A
RBB
|)
CP 5
C/S
Endpoint with Standard
C = F
A
S
CP 6
C/S/Rb
Endpoint with Standard
C = F
(A
S
- A
RB
)
CP 7
C/S/Sb
Endpoint with Standard
C = F
|A
S
- A
SB
|
CP 8
C/S/SbRb
Endpoint with Standard
C = F
(|A
S
- A
SB
| - |A
RB
- A
RBB
|)
CP 9
FTK/F/Rb
Fixed Time Kinetic with Factor
C = F
( |A
S,0
- A
S,1
| - |A
RB,0
- A
RB,1
| )
CP 10
FTK/S/Rb
Fixed Time Kinetic with Standard
C = F
( |A
S,0
- A
S,1
| - |A
RB,0
- A
RB,1
| )
CP 11
KIN/F/Rb
Kinetic with Factor
C = F
(
A
S,Minit
-
A
RB,Minit
)
CP 12
KIN/S/Rb
Kinetic with Standard
C = F
(
A
S,Minit
-
A
RB,Minit
)
CP 13
TRANSM.
Transmission in %
CP 14
C/F DELTA
Endpoint with Factor
C = F
(
A
S2-Sb2
-
A
S1-Sb1
)
CP 15
C/F 3 WL
Measurement with 3 Wavelengths
C = 168 * A
415nm
– 84 * A
380nm
– 84 *
A
450nm
CP 16
DELTA R1R2
Diff. measurement of two reagents
C =
A
S
Explanations:
CP-No. ........................ Number of the calculation procedure (chapter 7 - METHOD EDITOR)
Characteristic .............. Name of the calculation procedure (chapter 13.1 - BASIC METHOD)
Calculation formula ..... Calculation basis of basic method