![Rice Lake SCT20-AN Скачать руководство пользователя страница 11](http://html1.mh-extra.com/html/rice-lake/sct20-an/sct20-an_technical-manual_1454277011.webp)
Introduction
© Rice Lake Weighing Systems
●
All Rights Reserved
7
1.9
Key Functions
Figure 1-5. SCT Weight Transmitter Display and Keys
Table 1-3. Key Definitions
KEY
Short press in weigh mode
Long press
(3 sec) in weigh
mode
Function within Setup Menus
Escape
—
Zero Setting
Escape from a parameter or return to previous menu or operation mode.
Scroll/ Backspace
Captures Tare
Gross
Net
Removes Tare
Net
Gross
Move to the previous parameter in a level or scroll to the next digit in a
parameter value.
Next/ Data Entry
Save to alibi memory (if present) mV load cell test Move to the next parameter in a level or increment a value in a parameter.
Enter
Setting setpoints and hysteresis
—
Move to next level of configuration or select and edit a parameter.
and
Enter configuration for setting general parameters (press and hold
then press
to enter setup menu).
and
Setting preset tare (press and hold
then press
to enter setup menu.
Note
The LEDs are activated in sequence to indicate that a setting and not a weight is being viewed.
After pressing the first digit flashes and can be edited.
For numeric entries, press to select desired digit, then press to increment the digit.
1.10 Instrument Commissioning
1. Plug power cord into an outlet to power on the indicator. The display shows in sequence:
•
followed by the software code (
)
•
followed by the software version (
)
•
followed by the hardware code (
)
• The serial number (
)
2. Check that the display shows the weight and that there is an increase in weight when loading the load cells. If there is
not, check and verify the connections and correct positioning of the load cells.
3. Confirm that the SCT Weight Transmitter has been theoretically calibrated, the load cell’s rated data is already entered
and that the plant system identification tag is present on the instrument and on the cover.
Note
If the instrument has not been calibrated, follow the calibration instructions in
before
proceeding to next step.
4. Reset to zero. See
5. Check the calibration with test weights and correct the indicated weight if necessary. See
.