SCT-1100 Weight Transmitter
5. Press
. The weigh information displays.
6. Press
or
to view the weigh information.
•
–
is the scale number (from 1 to 4)
•
–
is the unit of measure (
,
,
)
•
momentarily displays and then the gross weight value
•
or
(manual tare) momentarily displays and then the tare weight value
7. Press to return to the weigh mode.
If the alibi memory is empty and
is pressed,
displays momentarily and the instrument returns to the
weigh mode.
If the entered ID is not valid,
displays and the instrument returns to the weigh mode.
Clearing the Alibi Memory
The alibi memory can be cleared directly on the instrument in the
parameter.
1. Turn on the instrument and press
while the firmware version displays.
displays.
2. Press
until
displays. Press
to enter the menu.
3. Press
or
until
displays. Press
.
displays.
4. Press
to clear the alibi memory or any other key to cancel.
•
displays if the operation is successful
•
displays if the memory was not successfully cleared (repeat procedure)
5. Press to return to the weigh mode.
It is not possible to clear an individual weigh record.
Serial Commands in Alibi Memory Mode
In addition to the serial commands described in
, the following serial commands are
available in the alibi memory mode. These commands are ignored if the program is not in alibi memory mode.
Weigh Storage
Command:
[II]PID<CRLF>
or
<ESC>[II]PID<STX>
and
[II]PIDD<CRLF>
or
<ESC>[II]PIDD<STX>
In which:
• [II] is the RS-485 address
• <ESC> is the 27 ASCII decimal character
• <STX> is the 2 ASCII decimal character
Instrument response:
• to the
[II]PID<CRLF>
command:
[II]PIDSS,B,LLLLLLLLLLUU,YYTTTTTTTTTTUU,(ID | NO) <CRLF>
• to the
<ESC>[II]PID<STX>
command:
<ESC>[II]PIDSS,B,LLLLLLLLLLUU,YYTTTTTTTTTTUU,(ID | NO)<STX>
• to the
[II]PIDD<CRLF>
command:
[II]PIDDSS,B,LLLLLLLLLLUU,YYTTTTTTTTTTUU,(ID | NO),(dd/mm/
yybbhh:mm:ss|NO DATE TIME)<CRLF>
• to the
<ESC>[II]PIDD<STX>
command:
<ESC>[II]PIDDSS,B,LLLLLLLLLLUU,YYTTTTTTTTTTUU,(ID | NO),(dd/mm/
yybbhh:mm:ss|NO DATE TIME)<STX>
In which:
• [II] is the RS-485 address (only when transmitting in RS-485 mode)
• SS can be – OL (weight in overload) or UL (weight in underload) or ST (stable weight) or US (unstable weight) or TL
(TILT input closed).
• , is a comma character (ASCII decimal 44)
•
B
is the scale number (from 1 to 4)
•
LLLLLLLLLL
is the gross weight on 10 digits
•
UU
is the unit of measure
C
Note
C