682 Synergy Plus
112
Visit our website
16.6 Stream Format Tokens
Table 16-3. Stream Format Tokens
Format Identifier
Defined By
Description
<P[G | N | T]>
STRM.POS#
n
STRM.NEG#
n
Polarity – Specifies positive or negative polarity for the current or specified
(Gross/Net/Tare) weight on the source scale; Possible values are SPACE, NONE,
+ (for STR.POS#
n
) or – (for STR.NEG#
n)
<U[P | S | T]>
STRM.PRI#
n
STRM.SEC#
n
STRM.TER#
n
Units – Specifies primary, secondary or tertiary units for the current or specified weight
on the source scale
<M[G | N | T]>
STRM.GROSS#
n
STRM.NET#
n
STRM.TARE#
n
Mode – Specifies gross, net or tare weight for the current or specified weight on the
source scale
<S>
STRM.MOTION#
n
STRM.RANGE#
n
STRM.OK#
n
STRM.INVALID#
n
STRM.ZERO#
n
Status for the source scale – Default values and meanings for each status:
• STR.MOTION#
n
M
In motion
• STR.RANGE#
n
O
Out of range
• STR.OK#
n
<space>
OK
• STR.INVALID#
n
I
Invalid
• STR.ZERO#
n
Z
COZ
<UID>
UID
Unit ID number – Specifies the unit identification number as an alphanumeric value up
to 6 characters long
<B [–]n,…>
See descriptions below
Bit fields. Comma-separated sequence of bit field specifiers; Must be exactly 8-bits;
minus sign ([–]) inverts the bit
B0
--
Always 0
B1
--
Always 1
B2
Configuration
=1 if even parity
B3
Dynamic
=1 if MODE=NET
B4
Dynamic
=1 if COZ
B5
Dynamic
=1 if standstill
B6
Dynamic
=1 if gross negative
B7
Dynamic
=1 if out of range
B8
Dynamic
=1 if secondary/tertiary
B9
Dynamic
=1 if tare in system
B10
Dynamic
=1 if tare is keyed
B11
Dynamic
=00 if MODE=GROSS
=01 if MODE=NET
=10 if MODE=TARE
=11 (not used)
B12
Dynamic
=00 if UNITS=PRIMARY
=01 if UNITS=SECONDARY
=10 if UNITS=TERTIARY
=11 (not used)
B13
Configuration
=00 (not used)
=01 if current DSPDIV=1
=10 if current DSPDIV=2
=11 if current DSPDIV=5
B14
Configuration
=00 (not used)
=01 if primary DSPDIV=1
=10 if primary DSPDIV=2
=11 if primary DSPDIV=5
B15
Configuration
=00 (not used)
=01 if secondary DSPDIV=1
=10 if secondary DSPDIV=2
=11 if secondary DSPDIV=5
B16
Configuration
=00 (not used)
=01 if tertiary DSPDIV=1
=10 if tertiary DSPDIV=2
=11 if tertiary DSPDIV=5