MSI Dyna-Clamp Tension Meter

12

Visit our website 

www.RiceLake.com

4.1.1 Error Codes

The ScaleCore processor of the Dyna-Clamp detects errors and generates error codes to aid in troubleshooting. 

4.2 Battery Replacement 

1. Unscrew the battery cover on the side of the handle.
2. Insert two AA batteries positive end first (can be replaced with rechargeable NiMH cells).
3. Reinstall the battery cover securely.

 

Figure 4-1. Battery Installation

4.2.1 Disposable Batteries

Low Battery

 displays when the battery is getting low. When 

Low Battery

 starts to blink, the batteries are close to being drained. 

For maximum life, use the batteries until the system shuts off.

4.2.2 Rechargeable Batteries

When using Nickel-Metal-Hydride (NiMH) cells, it is recommended that the cells are recharged immediately after 

Low Battery

 

starts to blink. Do not allow the batteries to discharge completely, it may compromise the recharge life of the battery.
NiMH Cells have a lower capacity than Alkaline. Having two sets of NiMH batteries is recommended, so one set can be 

charging while the other is in use.

The use of NiCad batteries is not recommended.
Remove the batteries prior to storing to prevent possible battery corrosion if the Dyna-Clamp is not going to be 

used for an extended period of time.

Error Code

Description



LC Disabled – A load cell was not enabled



Not Calibrated – System has not been calibrated





Overload – Tension exceeds set ca8 d; 

load Cell damaged or disconnected





Underload – Tension is more than 10% negative; 

load cell damaged or disconnected

Table 4-2.  Error Codes

Part No.

Description

Qty

191147

Battery Cap Assembly

1

16431

AA Cell 1.5 V Alkaline Battery

2

Table 4-3.  Battery Parts List

AA Batteries

Battery Cover

IMPORTANT

Содержание MSI Dyna-Clamp

Страница 1: ...MSI Dyna Clamp Tension Meter Operation Manual May 19 2020 PN 193853 Rev A...

Страница 2: ...marks of their respective companies All information contained within this publication is to the best of our knowledge complete and accurate at the time of publication Rice Lake Weighing Systems reserv...

Страница 3: ...sion Measurement 4 2 7 Average Function 5 2 7 1 Average Procedure 5 2 7 2 Clearing Average 5 2 8 Peak Function 5 2 8 1 Clearing Peak 5 2 9 Safety Strap 5 2 10 Carry Case Option 5 3 0 Configuration 6 3...

Страница 4: ...na Clamp Tension Meter ii Visit our website www RiceLake com Rice Lake continually offers web based video training on a growing selection of product related topics at no cost Visit www ricelake com we...

Страница 5: ...nderstood Failure to follow the instructions or heed the warnings could result in injury or death Contact any Rice Lake Weighing Systems dealer for replacement manuals Failure to heed could result in...

Страница 6: ...Battery displays when approximately 10 of battery life remains and blinks when automatic shutdown is imminent 6 Primary display with six 0 7 18 mm numeric digits displays live wire rope tension and is...

Страница 7: ...n extended period of time 2 2 Product Dimensions Figure 2 1 Product Diagram 2 3 Power Press to turn on the Dyna Clamp The startup sequence displays all LCD segments followed briefly by the software ve...

Страница 8: ...y grease to the drive rod threads to maintain proper lubrication 1 Loosen the drive rod 2 Zero the Dyna Clamp Section 2 5 Figure 2 2 Dyna Clamp On Wire Rope 3 Position the Dyna Clamp on the wire rope...

Страница 9: ...Clearing Average Follow this procedure when the running average needs to be cleared 1 Position the Dyna Clamp on the wire rope and tighten slightly so a little positive tension is measured by the dev...

Страница 10: ...ls through the available wire rope types and the average and peak values in the secondary top display area Section 3 3 4 on page 7 ActFun performs action based on the active function that is displayed...

Страница 11: ...counting from to followed by the c cal value displaying in the secondary display area The test can be single stepped 1 Press F key set to then immediately press to stop the auto scroll 2 Press to scro...

Страница 12: ...gh setpoint value then the setpoint 2 LED turns on If the tension measurement is within the specified range then neither setpoint LED is on Use the following steps to set the Range setpoint values 1 P...

Страница 13: ...s in the secondary top display area 4 Press to scroll through the available filter settings 5 With desired filter displayed press to accept displays 6 Press to save and return to tension mode 3 7 Unit...

Страница 14: ...if it is in the setup menu The auto off parameter Section 3 4 on page 8 also conserves battery life by turning the unit off if no activity is sensed for a set time The standard mode is recommended for...

Страница 15: ...before load is lifted Zero the Dyna Clamp with no load attached lbf kgf kN units causing confusion Select proper units Requires recalibration Requires authorized service Display reading not stable Exc...

Страница 16: ...ls are recharged immediately after Low Battery starts to blink Do not allow the batteries to discharge completely it may compromise the recharge life of the battery NiMH Cells have a lower capacity th...

Страница 17: ...4 Bobbin Replacement 1 Remove the bolt securing the wire rope bobbin to the frame and discard old bobbin 2 Align new wire rope bobbin with hole in frame 3 Torque bolt to 30 ft lb 40 N m with blue Loc...

Страница 18: ...c digits Secondary Display top Eight 0 5 13 mm numeric digits Function Keys Programmable Keys Two user programmable F1 F2 TEST UNIT PRINT SCROLL ACTFUN Permanent Keys Two fixed POWER ZERO Setpoints Tw...

Страница 19: ......

Страница 20: ...anada Mexico 800 321 6703 International 715 234 9171 Europe 31 0 26 472 1319 Rice Lake Weighing Systems is an ISO 9001 registered company Rice Lake Weighing Systems Specifications subject to change wi...

Отзывы: