5.6 Calibration Setup Menu
The Calibration Setup Menu contains two additional items beyond Calibration:
• Standard menu
• Auto Zero Maintenance menu (
Auto0
)
In addition, more menus will appear that are transferred from the main setup menu when Legal-for-Trade settings are used.
5.6.1 Standard Menu
Use the following steps to set up a Legal for Trade standard settings.
1. Remove the hex seal screw from the MSI-4260M using the steps from
and
displays.
2. Press
twice to scroll so
displays.
3. Press
to enter the Calibration setup menu.
displays.
4. Press
to enter the standard menu. The current standard setting displays.
5. Press
to scroll to the desired standard.
6. Press
to set the standard.
displays.
7. Press
twice to exit setup and store all changes.
displays.
Selection
Description
Industrial (indu5)
This is the most common setting for the MSI-4260M; With the Industrial standard, you have full range zero, access to units
switching, filters, and peak hold
Handbook 44 (Hb-44) Sets the scale to enable only approved features per the NTEP HB-44 rules and regulations; Access is denied to Peak Hold,
and the zero range may be limited; The Filter menu is moved to the Cal Setup Menu, so filters are only accessible through the
Cal Seal
R-76 (r-76)
Sets the scale to enable only approved features per OIML R-76; Only kg weight units are available; The zero range is limited to
4% (-1–3% relative to Calibrate zero); Net/Gross function is temporary; When Net weight is established, pushing an F key set
for Net/Gross will cause a maximum 5 second display of the Gross weight; Tare must be cleared to display Gross weight
constantly; Other metrological aspects are changed to meet R-76 requirements
One Unit (1unit)
The one unit Standard is exactly the same as Industrial, except units switching is inhibited; This is useful for Metric only
countries; Another use of the One Unit standard is to allow the scale to be calibrated in units other than lb or kg, since
conversions are eliminated; Contact Rice Lake Weighing Systems for more information on the Standards settings
Table 5-3. Standard Menu Selections