IQ plus 350 Installation Manual
TEST Menu
Function
Description
A/DTST
Display A/D test
Press and hold Enter key to display raw count from A/D converter.
ADOFFS
A/D offset calibration (–0.5 mv/V)
Read Caution! statement on page 41 before using this function.
Press setup switch and Enter key at
the same time to perform offset calibration.
ADGAIN
A/D gain calibration (+4.5 mv/V)
Read Caution! statement on page 41 before using this function.
Press setup switch and Enter key at
the same time to perform gain calibration.
D/I 1
Display digital input 1
Press and hold Enter key to show status of DIGIN1 (DI1=HI or DI1=LO).
D/I 2
Display digital input 2
Press and hold Enter key to show status of DIGIN2 (DI2=HI or DI2=LO).
AOUT=0
Set analog output to zero
Press and hold Enter key to set analog output to its zero value.
AOUTFS
Set analog output to full scale
Press and hold Enter key to set analog output to its full scale value.
DEFLT
Default parameters
Press setup switch and Enter key at the same time to reset configuration and calibration parameters to factory
default values.
XMT U
Transmit “U”
Press and hold Enter key to send ASCII “U” characters (decimal 85) from the serial port.
ECHO R
Echo received characters
Press and hold Enter key to view characters received at serial port.
NOTE:
IQ plus 350 display shows lower-case characters as blanks.
Table 7-7. Test Menu Functions