4.0
Maintenance
This section provides an overview of BenchMark LP Low-Profile Bench Scale maintenance information.
4.1 Troubleshooting
For troubleshooting details, see the following information:
Periodic Maintenance
The space beneath the platform must be periodically cleaned to prevent debris buildup.
Do not attempt to use scales with non-hermetically-sealed load cells in washdown applications. Water damage
is a common cause of failure in non-hermetically-sealed load cells.
Use care with high pressure steam washdowns for hermetically-sealed load cells. Steam will not damage the
load cells, but the elevated temperatures may cause incorrect readings until the unit cools to room temperature.
Problem
Possible Cause
Solution
The system does not
operate; There is nothing
on the display
The power is disconnected
Check and reconnect the power
The indicator fuse is blown
Replace the fuse; Check for the cause
The interface cable is cut or disconnected
Repair the cable
Signal leads are incorrectly installed at the indicator Install the leads according to the indicator installation manual
The display stays at zero The indicator is faulty
Service the indicator
The load cell connections are faulty
Check the cable connections in the summing board and at the indicator
Weights display erratically There is vibration near the scale
Remove the source of vibration, or move the scale
The platform is not level within 1/4''
Level the scale by adjusting the feet, or by shimming (if necessary)
The load cell or cable has water damage
Replace the load cell or cable
There is debris under the load cells or platform
Clean under the load cells and platform
The indicator is faulty
Use a simulator to test the indicator for stability, then service the indicator
Weights are consistently
high or consistently low
The indicator is not properly adjusted to zero
Zero the indicator according to the indicator manual
The platform is binding
Obtain adequate clearance so the platform has free movement
The indicator is not calibrated
Calibrate according to the indicator manual; Additionally, see
of this manual
The load cells are faulty
Test the load cells, replace if necessary
The feet are touching the deck underside
Adjust the feet downward to provide adequate clearance
Table 4-1. Troubleshooting Guide