482-AG Livestock Digital Weight Indicator
18
Visit our website
3.1 Configuration Menu
Figure 3-1. Configuration Menu
Parameter
Settings
Description
GRADS
4000
1 – 100000
Graduations – specifies the number of full scale graduations; the value entered must be in the range of 1–100,000 and
should be consistent with legal requirements and environmental limits on system resolution
ZTRKBN
0.6 dd
0.0 – 3.0dd
Zero track band – automatically zeros the scale when within the range specified, as long as the input is within the configured
zero range (ZRANGE parameter); selections are ± display divisions; maximum legal value varies depending on local
regulations
ZRANGE
2%
1%–100%
Zero range – selects the range within which the scale can be zeroed; the 2% selection is ± 2% around the
calibrated zero point, for a total range of 4%; indicator must be at standstill to zero the scale
INIZR
0%
0–100%
Initial zero range at power up; see Error messages HINOFF and LINOFF in
; if on power up, the
weight value is between the ±% range specified of Calibrated Zero, the indicator will zero off that weight; if outside that
range, the indicator will display HINOFF or LINOFF, depending on whether the initial reading is above or below the range
NOTE: If not set at zero, hopper/scale MUST be empty when powered up.
MOTBAN
1.0d
0.0d–100d
Motion band – sets the level, in display divisions, at which scale motion is detected; if motion is not detected, the standstill
symbol lights; some operations, including print, tare and zero, require the scale to be at standstill; maximum legal value
varies depending on local regulations
NOTE: Setting the MOTBAN setting to 0.0 disables motion sensing and the scale will always be
considered stable.
OVRLOA
FS+2%
FS+1D
FS+9D
FS
Overload – determines the point at which the display blanks and an out-of-range error message is displayed; maximum
legal value varies depending on local regulations
DSPRAT
0.1 SEC
0.25 SEC
0.5 SEC
0.75 SEC
1 SEC
NONE
Display rate – sets the update rate for displayed values; Values are in seconds (SEC)
No display filtering – the display is updated at up to the A/D sample rate
SMPRAT
5HZ
10HZ
20HZ
40HZ
A/D Sample rate – selects measurement rate, in samples per second, of the analog-to-digital converter; lower sample rate
values provide greater signal noise immunity
DIGFIL
ROLFIL
OFF
ADPFIL
The digital filter setting determines which if any digital filter option is used
• ROLFIL - Rolling Average Filter
• OFF - Disables digital filter
• ADPFIL - Adaptive Filter
ADPFIL
--
Adaptive Filter parameter settings, see
ROLFIL
--
Rolling Average Filter parameter settings, see
DSPMOD
--
Display Mode parameter settings, see
Table 3-2. Configuration Menu Parameters
CONFIG
FORMAT
CALIBR
PROGRM
PASWRD
TEST
GRADS
ZTRKBN
ZRANGE
MOTBAN
OVRLOA
DSPRAT
SMPRAT
ROLFIL
DIGFIL
ADPFIL
DSPMOD
INIZR