Testing and Care
Page 24 of 36
Safe Place Transmitters User Guide
0510-1121-G
Operation
Test the operation of the transmitter using
the Transmitter Tester (PN 9450-0050). The
transmitter tester will detect whether or not a
transmitter is emitting a signal, but cannot
indicate the strength of the signal
1. Place the transmitter tester directly on
the transmitter.
2. Press and hold the button on the left
side of the transmitter tester.
3. The device beeps once when you
initially press the button.
4. While holding the button in, the indicator
light flashes and a tone sounds once
per second, wait for at least 3 flashes of
the indicator light and 3 tones from the
transmitter tester to verify that the
transmitter is functioning correctly.
5. If the transmitter fails this test, check the
battery in the tester. The transmitter
tester requires a 9V battery.
6. Retest the transmitter. If it fails a second
time, DO NOT USE THE
TRANSMITTER and contact RF
Technologies Technical Support Team
at (800)669-9946 or (262) 790-1771.
Signal
Strength
1. Physically enter an Exit Zone with the
transmitter and walk through the door.
2. An alarm must occur when the
transmitter is within 4-feet of the
monitored door. If applicable, the door
also locks.
3. If an alarm does not occur, DO NOT
USE THE TRANSMITTER and contact
RF Technologies Technical Support
Team at (800)669-9946 or (262) 790-
1771.