Rexel DT-326 Скачать руководство пользователя страница 1

Air Quality CO2 Monitor 

DT-326

1. Description

The Carbon Dioxide (CO

2

) Monitor is designed for air quality control and health 

control by measuring Carbon Dioxide (CO

2

) level in areas where it needs to be 

monitored, such as factories, schools, offices, transportation lines and so on.The 

measured CO

2

 value in ppm (parts-per-million) will be displayed on the LCD with 

three acoustic alarm sounds when then CO

2

 level is exceeding the define level.

2. Feature

• NDIR (non-dispersive infrared) CO

2

 sensor

• Checks for Carbon Dioxide (CO

2

) concentrations

• Indoor Air Quality displayed in ppm with Good (0 to 800ppm), Normal (800 

to1500ppm), Poor (>1500ppm) indication

• Easy to read, large dual display 

• Simultaneously displays  (CO

2

) concentrations

Air Temperature and Humidity

• Adjustable carbon dioxide refresh frequency.

• Adjustable carbon dioxide alarm function in the range of 0-5000 ppm.

• Air quality indicator function. 

• low battery indication.

3. Specifications

Range

Accuracy

Resolution

CO

2

0 to 5000ppm ±75 ppm ±5% of reading

1 ppm

Temperature

-10 to 50

o

C

±1

o

C

0.1

o

C/

o

F

Humidity

0 to 100 %RH

±3.5 %RH (20%RH-80%RH)

±5%RH (0%RH-20%RH/ 

80%RH-100%RH)

0.1 %RH

Power supply 

1.5V AA Battery * 2

Charging

Micro USB DC 5V 1A

Working temperature -10 to 50

o

C

Storage temperature -10 to 60

o

C

4.Names and function of parts LCD display 

1. LCD

2. °C/°F Switch Function Button (UP Button)

3. Power Button

4. SET Button(UP Button)

5. Instrument Hang Buckle Point

6. CO

2

 sensor

7. Temperature and humidity sensor

8. Battery Cover

9. Power supply Micro USB DC 5V 1A

5. LCD Introduction (Below is the Picture of LCD)

1. CO

2

 display value

2. Temperature display value

3. Humidity display value

4. Battery power icon

5. Temperature unit : Celsius 

     and Fahrenheit 

6. Humidity unit : %RH

7. CO

2

 unit : PPM

8. Color screen printing and air 

    quality indication

9. Temperature icon

10.Humidity icon

6. Operation Instructions

Key function

In the measurement mode:

Press the °C/°F switch function button briefly will switch between Celsius and 

Fahrenheit;

Press the SET button briefly to enter the setting mode;

Long press the power button to shut down;

In the setting mode:

Press the °C/°F switch function button briefly to increase the corresponding 

(flashing) value;

Press the SET button briefly to decrease the corresponding (flashing) value;

Press the power button briefly to switch between alarm value setting and CO

refresh frequency setting;

Long press the SET button to save the set value and exit the setup mode;

Long press the power button to shut down.

·Start measuring

Install 2 x AA 1.5V batteries; Power supply Micro USB DC 5V 1A.

Press the power button to start the instrument;

After starting up,CO

2

 updates the data according to the set frequency;

The temperature and humidity update the real-time data once every second;

Air environment indication function:

Range(PPM)

Grid

Color

Grade

<800

1

 

GOOD

800~1500

2

 

NORMAL

>1500

3

 

POOR

The analog bar in the middle of the liquid crystal will indicate the current 

environmental quality according to the CO

2

 concentration.

Setting mode

Alarm value setting mode

After entering the setting mode, the alarm value setting mode will be entered 

firstly

Alarm value setting: PPM is displayed, CO

2

 value flashes, and alarm value unit is 

ppm. When  press the “

” or “

”  button , the alarm value will be increase by 

10ppm or decrease by 10ppm.the default value is 1500;

Alarm value setting mode diagram is as follows:

CO

2

 sampling frequency setting mode

In the alarm value mode,press the power button briefly to enter CO

2

 sampling 

frequency setting mode. 

9