Parameters for the installation location
Ambient temperature:
5°C to 40°C
If the ambient temperature falls below or exceeds the limits
specified, the electrical and mechanical components may be dam-
aged or the performance data changed to an unknown extent..
Humidity :
Maximum relative humidity 80% at temperatures up to 31°C,
decreasing in a straight line to 50% relative humidity at 40°C.
At higher humidity, the electrical and mechanical components
may be damaged and performance data changed to an unknown
extent.
Site altitude:
max. 2000 m above sea level
Installation
Place the BB 51 jaw crusher on an even and stable surface, eg. a labo-
ratory bench.
Electrical connection
•
Consult the type plate for the voltage and frequency of the BB
51.
•
Ensure that the values are in agreement with the mains power
supply.
•
Connect the BB 51 to the mains using the cable supplied.
Failure to observe the values on the type plate can result in dam-
age to electronic or mechanical components.
17.12.1998
©
Retsch GmbH
9
Doc. No. GB 98.056.9999