Cleaning, Wear and Maintenance
36
10 Cleaning, Wear and Maintenance
10.1 Cleaning
WARNING
W8.0003
Danger to life through electric shock
Cleaning with water
An electric shock can cause burns, cardiac arrhythmia, respiratory arrest,
as well as cardiac arrest.
•
The power cable must be unplugged before cleaning the device.
•
Use a cloth dampened with water for cleaning.
•
Do not clean the device under running water!
NOTICE
N15.0009
Damage to the housing and device
Use of organic solvents
Organic solvents may damage plastic parts and the coating.
•
The use of organic solvents is not permitted.
Clean the housing of the device with a damp cloth and if necessary, with a household
cleaning agent. Pay attention that no water or cleaning agent enters the interior of the
device.
10.1.1 Cleaning of Test Sieves
Test sieves are measuring instruments and should be treated with due care before, during and
after the sieving process. It is recommended to clean new test sieves before the first use from
possible preservative residues with ethanol or isopropanol and to store them in a dry, dust-free
place when unused.
Before cleaning or drying the test sieves, the O-rings have to be removed. Before using and
after the cleaning the test sieves should be visibly inspected for possible damages and
impurities.
Near-mesh or clamped particles can be often removed dry after the sieving process by slightly
tapping the test sieve upside down with the sieve frame on a table. For test sieves with mesh
sizes > 500 µm a fine hair brush can be used to sweep over the outer side of the mesh fabric.
10.1.1.1
Cleaning of Test Sieves with Mesh Sizes > 500 µm
Coarse mesh fabrics with mesh sizes > 500 µm can be cleaned dry or wet easily and effectively
with a hand brush with plastic bristles (at not too high applied pressure).
10.1.1.2
Cleaning of Test Sieves with Mesh Sizes < 500 µm
Test sieves with mesh sizes < 500 µm should generally only be cleaned in an ultrasonic
cleaning-bath. As cleaning agent, water together with a standard surfactant is recommended.
The cleaning in the ultrasonic bath usually takes two to three minutes. After that the test sieves
Содержание AS200 tap
Страница 2: ...2 Copyright Copyright by Retsch GmbH Retsch Allee 1 5 42781 Haan Germany ...
Страница 5: ...Notes on the Manual 5 ...
Страница 10: ...Notes on the Manual 10 Service address ...
Страница 15: ...Technical Data 15 3 11 Drive Horizontal circular motions 280 rpm Taps 150 min 1 ...
Страница 21: ...Packaging Transport and Installation 21 Fig 5 Minimum distance between housing and hoist 85 cm 34 inches ...
Страница 44: ...Index 44 Wet sieving 14 Width 13 base 13 Workplace related emission level 12 Y Year of production 18 ...
Страница 45: ...45 ...
Страница 46: ...46 ...
Страница 47: ...47 ...
Страница 48: ...Copyright Copyright by Retsch GmbH Retsch Allee 1 5 42781 Haan Germany ...