First Commissioning
24
When switching on again, press the
(
H3
) and
(
H4
) buttons of the time display
simultaneously. "P2" appears in the time display.
Press the
button (
H1
) to enter the standby mode.
To return to the "P1" programme mode, repeat this process. "P1" appears in the time
display.
Fig. 7:
Programm modes "P1" (top) and "P2" (bottom) of the device
5.2 Inserting the Test Sieves
CAUTION
C6.0012
Contusions and bruises
Overturning of the sieve stack
The sieve stack can overturn and cause personal injury.
•
Only operate the device with securely clamped sieve stack.
To insert the sieve stack more easily, the tapping arm (
A
) can be secured in an elevated
position with the locking bolt (
F
).
Pull the locking bolt (
F
) forward to the stop.
Lift up the tapping arm (
A
), so that the bore of the tapping arm is at the same level with the
locking bolt.