Packaging, Transport and Installation
17
4.4 Conditions for the Installation Site
Installation height: max. 2
000 m above sea level
Ambient temperature: 5 °C
– 40 °C
NOTICE
N8.0021
Ambient temperature
Temperatures outside the permitted range
Electronic and mechanical components may be damaged.
The performance data alter to an unknown extent.
•
Do not exceed or fall below the permitted temperature range (5 °C to
40 °C ambient temperature) of the device.
Maximum relative humidity < 80 % (at ambient temperatures
≤ 31 °C)
For ambient temperatures U
T
between 31 °C and 40 °C, the maximum relative humidity value L
F
linearly decreases according to L
F
=
–(U
T
– 55) / 0.3:
Ambient temperature
Max. rel. humidity
≤ 31 °C
80 %
33 °C
73.3 %
35 °C
66.7 %
37 °C
60 %
39 °C
53.3 %
40 °C
50 %
NOTICE
N9.0015
Humidity
High relative humidity
Electronic and mechanical components may be damaged.
The performance data alter to an unknown extent.
•
The relative humidity in the vicinity of the device should be kept as
low as possible.
4.5 Electrical Connection
WARNING
W3.0015
Danger to life through electric shock
Connection to sockets without protective earth
An electric shock can cause burns, cardiac arrhythmia, respiratory arrest,
as well as cardiac arrest.
•
The device may only be operated on sockets with protective earth
(PE).
Содержание AS200 tap
Страница 2: ...2 Copyright Copyright by Retsch GmbH Retsch Allee 1 5 42781 Haan Germany ...
Страница 5: ...Notes on the Manual 5 ...
Страница 10: ...Notes on the Manual 10 Service address ...
Страница 15: ...Technical Data 15 3 11 Drive Horizontal circular motions 280 rpm Taps 150 min 1 ...
Страница 21: ...Packaging Transport and Installation 21 Fig 5 Minimum distance between housing and hoist 85 cm 34 inches ...
Страница 44: ...Index 44 Wet sieving 14 Width 13 base 13 Workplace related emission level 12 Y Year of production 18 ...
Страница 45: ...45 ...
Страница 46: ...46 ...
Страница 47: ...47 ...
Страница 48: ...Copyright Copyright by Retsch GmbH Retsch Allee 1 5 42781 Haan Germany ...