Technical Data
12
3
Technical Data
3.1 Degree of Protection
IP50
3.2 Emissions
CAUTION
C4.0011
Possibility of acoustic signals not being heard
Loud sieving noises
Possible acoustic alarms and voice communication might not be heard.
•
Consider the volume of the sieving noise in relation to other acoustic
signals in the work environment. Additional visual signals may be
used.
CAUTION
C5.0025
Hearing damage
A high sound level may be generated depending on the type of material, the
number of sieves, the sieving aid used and the duration of the sieving
Excessive noise in terms of level and duration can cause impairments or
permanent damage to hearing.
•
Ensure suitable noise protection measures are taken or wear ear
protection.
Sound parameters:
The sound parameters are also influenced by the number of test sieves and the properties of
the sample material.
Example 1:
Number of test sieves:
5
Feed material:
Quartz sand (< 1 mm)
Soundproof hood:
no
At these operating conditions, the workplace related equivalent continuous sound level
L
eq
= 83 dB(A).
Example 2:
Number of test sieves:
5
Feed material:
Quartz sand (< 1 mm)
Soundproof hood*:
yes
*of Retsch GmbH
At these operating conditions, the workplace related equivalent continuous sound level
L
eq
= 65 dB(A).
Содержание AS200 tap
Страница 2: ...2 Copyright Copyright by Retsch GmbH Retsch Allee 1 5 42781 Haan Germany ...
Страница 5: ...Notes on the Manual 5 ...
Страница 10: ...Notes on the Manual 10 Service address ...
Страница 15: ...Technical Data 15 3 11 Drive Horizontal circular motions 280 rpm Taps 150 min 1 ...
Страница 21: ...Packaging Transport and Installation 21 Fig 5 Minimum distance between housing and hoist 85 cm 34 inches ...
Страница 44: ...Index 44 Wet sieving 14 Width 13 base 13 Workplace related emission level 12 Y Year of production 18 ...
Страница 45: ...45 ...
Страница 46: ...46 ...
Страница 47: ...47 ...
Страница 48: ...Copyright Copyright by Retsch GmbH Retsch Allee 1 5 42781 Haan Germany ...