7-7
7. Alarms
to match the alarm priority. Red background means
that the highest priority of the multiple alarm messages
is high while yellow background means that the highest
priority of the multiple alarm messages is medium. You
can view active alarms by selecting the alarm message
field.
: indicates that there is inactivated alarms for which the
alarm trigger condition has disappeared. Press this icon
to view the most recent inactivated alarms in the Alarm
list window. You can also clear the most recent alarms
with the [
Reset
] key.
: indicates that the alarm of a parameter is closed and
the alarm signal is in the ALARM OFF mode.
7.5 Alarm Volume Settings
Set Alarm Volume:
1. Select [
Alarm limit
] → [
Volume
].
2. Set [
Alarm Volume
]: X-10, with X being the
minimum alarm volume and 10 the maximum alarm
volume. If there are no currently active alarms, you
can also select one level. The system will emit a
low-priority alarm tone once based on the selected
alarm volume.
Set minimum alarm volume:
1. Select [
Setup
] → [
Maintain
] → Enter system
maintain password → [
System
] → [
Alarm
].
2. Set [
Minimum Alarm Volume
] to an appropriate
value.
WARNING
•
Do not rely exclusively on the audible alarm
system when using the ventilator.
•
Adjustment of alarm volume to a low level may
result in a hazard to the patient.
•
Always keep the patient under close surveillance.
Содержание iHope RS Series
Страница 1: ...iHope Ventilator User Manual RS Series...
Страница 10: ......
Страница 12: ......
Страница 26: ...1 14 1 Safety Information For your note...
Страница 28: ......
Страница 36: ...2 8 2 General Information For your note...
Страница 38: ......
Страница 58: ...3 20 3 Installations and Connections For your note...
Страница 60: ......
Страница 74: ......
Страница 88: ...5 14 5 System Settings For your note...
Страница 122: ...6 32 6 Start Ventilation For your note...
Страница 158: ......
Страница 168: ......
Страница 186: ......
Страница 200: ...10 14 10 Cleaning and Disinfection For your note...
Страница 202: ......
Страница 203: ...11 1 11 Specifications 11 Specifications 11 1 Physical Characteristic...
Страница 230: ...11 28 11 Specifications For your note...
Страница 232: ......
Страница 233: ...A 1 A Pneumatic Diagram A Pneumatic Diagram Figure A 1 Pneumatic diagram...
Страница 240: ...C 2 C Communications Interface For your note...
Страница 248: ...E 2 E Compliance For your note...