6-24
6. Start Ventilation
6.5.15 VS
(Volume Support Ventilation Mode)
VS achieves the target volume by regulating the
pressure applied following an initial pressure ramp-up.
VS delivers time-cycled mandatory breaths and
pressure-supported spontaneous breaths.
The mode is only available for single-limb non-invasive
ventilation .
P-max
P-min
EPAP
1/f
Mandatory (timed)
breath
Rise
Patient-triggered (Spont)
spontaneous breath
lnspiration
Time
Figure 6-18.VS waveform
Table 6-15.VS settings
Parameter
Description
O
2
%
Adjustment of oxygen concentration.
VT
Tidal volume of controlled breaths.
f
Breathing frequency.
T-insp.
Time of inspiration.
Rise
Rise time of inspiratory pressure.
EPAP
Positive end-expiratory pressure.
P-min
Minimum set IPAP.
P-max
Max set IPAP.
I-trigger
Inspiration trigger level.
E-cycle
Expiration trigger level.
6.5.16 PPS
(Proportional Pressure Support Ventilation Mode)
Содержание iHope RS Series
Страница 1: ...iHope Ventilator User Manual RS Series...
Страница 10: ......
Страница 12: ......
Страница 26: ...1 14 1 Safety Information For your note...
Страница 28: ......
Страница 36: ...2 8 2 General Information For your note...
Страница 38: ......
Страница 58: ...3 20 3 Installations and Connections For your note...
Страница 60: ......
Страница 74: ......
Страница 88: ...5 14 5 System Settings For your note...
Страница 122: ...6 32 6 Start Ventilation For your note...
Страница 158: ......
Страница 168: ......
Страница 186: ......
Страница 200: ...10 14 10 Cleaning and Disinfection For your note...
Страница 202: ......
Страница 203: ...11 1 11 Specifications 11 Specifications 11 1 Physical Characteristic...
Страница 230: ...11 28 11 Specifications For your note...
Страница 232: ......
Страница 233: ...A 1 A Pneumatic Diagram A Pneumatic Diagram Figure A 1 Pneumatic diagram...
Страница 240: ...C 2 C Communications Interface For your note...
Страница 248: ...E 2 E Compliance For your note...