Chapter 5
5-26
Esprit® Ventilator Service Manual © Respironics, Inc.
REF 580-1000-02 G
Repair
Filter Heater
WARNING:
If the ventilator has been operating, the filter heater sleeve may be hot.
1. Touch
Filter Heat
on the hardware screen to turn the filter heater on
(white background). Skip this step if the ventilator has been operating
in ventilation mode.
2. Wait 15 minutes. Skip this step if the ventilator has been operating in
ventilation mode.
3. Remove the exhalation filter and verify that it is very warm to the
touch.
4. Reinstall the filter and heater.
5. Return the hardware screen to its initialized state.
Blower
1. Connect a tube from the ventilator main outlet to the high flow port of
pneumatic calibration analyzer (Figure 5-16).
2. Set the analyzer’s peak/continuous function to
continuous
.
3. Set analyzer’s gas flow to measure 180 LPM air. Zero the analyzer.
4. Touch
Safety
to energize (white background) the safety solenoid.
5. Touch
Air
and set the flow to 165 LPM.
6. Check that the analyzer reads 148.5 to 181.5 LPM.
7. Touch
Blower
to turn off the blower (gray background).
8. Check that the blower stops and the analyzer reads 0 ± 0.1 LPM.