Chapter 5
5-24
Esprit® Ventilator Service Manual © Respironics, Inc.
REF 580-1000-02 G
Repair
2. Set analyzer to read LPM.
3. Set analyzer’s gas flow to air.
4. Touch
Safety
to energize (white background) the safety solenoid.
5. Check that the set air flow corresponds to measured flow:
6. Return the hardware screen to its initial state.
Check Valve 2
1. Connect the oxygen source to the ventilator.
2. Connect patient circuit to the ventilator, then block the wye.
3. Touch
Safety
to energize (white background) the safety solenoid.
4. Touch
Crossover
to de-energize (gray background) the crossover
solenoid.
5. Touch
Blower
to de-energize (gray background) the blower.
6. Touch
Exhalation
and set it to 2000 steps.
7. Touch
Oxygen
and set flow to 1 LPM.
8. Touch
Air
and set flow to 200 LPM.
9. Check that
Inhalation Pressure
display on the hardware screen reads
greater than 30 cmH
2
O.
10. Check that the
Air Flow
display on the hardware screen reads 0 ± 0.1
LPM.
11. Return the hardware screen to its initialized state.
Touch
Air
and Set the
Flow to:
Check that
Analyzer, Air Flow
and
Exhalation Flow
Displays
Read:
5 LPM
4.5 to 5.5 LPM
20 LPM
18 to 22 LPM
50 LPM
45 to 55 LPM
100 LPM
90 to 110 LPM
120 LPM
108 to 132 LPM
165 LPM
148.5 to 181.5 LPM
0 LPM
0 ± 0.1 LPM