Chapter 5
5-18
Esprit® Ventilator Service Manual © Respironics, Inc.
REF 580-1000-02 G
Repair
6. Touch
Crossover
to de-energize (gray background) the crossover
solenoid.
7. Touch
Oxygen
and set the flow to 180 LPM.
8. Check that the gauge reads 23 ± 1 PSI. If not, adjust the oxygen
regulator pressure:
a.
Loosen the locking nut.
b.
Turn the adjusting shaft clockwise to increase or counterclockwise
to decrease so the gauge reads 23 ± 1 PSI, then tighten the
locking nut.
c.
Touch
Oxygen
and set the flow to 0 LPM and then back to 180
LPM.
d.
Check that the gauge still reads 23 ± 1 PSI. Repeat steps
a
to
d
if
necessary.
9. Disconnect the oxygen source from the ventilator.
10. Return the hardware screen to its initialized state.
11. Remove the gauge and reconnect tubing to the safety valve solenoid
tubing connection.
Air Valve
1. Disconnect the bacteria filter, patient circuit tube, and oxygen sensor
tee from the ventilator inspiratory outlet.
2. Touch
User Config
and note the altitude.
3. Touch
Hardware
.
4. Touch
Safety
to energize (white background) the safety solenoid.
5. Touch
Air
and set the flow to 1 LPM.
6. Check that the
Air Position
display is between 215 and 525 steps.
7. Touch
Air
and set the flow to 180 LPM.