Appendix A
A-20
Esprit® Ventilator Service Manual © Respironics, Inc.
REF 580-1000-02 G
Patient Circuit
(Test 11, Step 1)
This step pressurizes the circuit with air to measure tubing compliance and
leak rate, and includes several time-dependent sub-tests.
Diagnostic code
3106
indicates a leak in the circuit that prevents the circuit
from being pressurized. Diagnostic code
3107
occurs if a leak rate is
excessive. Diagnostic code
3125
occurs if PT2 and PT3 disagree by >3
cmH
2
O. Diagnostic code
3128
occurs if calculated compliance is <0.5 or
>9.0 mL/cmH
2
O. Diagnostic code
3129
indicates a leak in the circuit that is
out of range.
Red = air flow
Blue = O
2
flow
Black = no flow