
Appendix A
A-18
Esprit® Ventilator Service Manual © Respironics, Inc.
REF 580-1000-02 G
Exhalation Valve
(Test 10, Step 1)
This step commands the exhalation valve to specific positions to verify
accurate control.
Diagnostic code
3149
indicates that the exhalation valve stepper motor open
position is out of range. Diagnostic code
3150
indicates that the exhalation
valve stepper motor midpoint position is out of range. Diagnostic code
3151
indicates that the exhalation valve stepper motor closed position is out of
range.
Red = air flow
Blue = O
2
flow
Black = no flow
Содержание ESPRIT
Страница 1: ...REF 580 1000 02 G ESPRIT Ventilator S e r v i c e M a n u a l ...
Страница 318: ...Appendix A A 24 Esprit Ventilator Service Manual Respironics Inc REF 580 1000 02 G This page is intentionally blank ...
Страница 320: ...Appendix B B 2 Esprit Ventilator Service Manual Respironics Inc REF 580 1000 02 G This page is intentionally blank ...