Crossover Circuit
(Test 5, Step 1)
This step pressurizes the circuit with air (oxygen is disconnected) to verify that
the crossover solenoid operates properly when energized.
Diagnostic code
3112
occurs if oxygen is not disconnected. Diagnostic code
3117
indicates a crossover circuit fault. Diagnostic code
3119
occurs if the
blower does not turn off.
Red = air flow
Blue = O
2
flow
Black = no flow