
Appendix A
A-10
Esprit® Ventilator Service Manual © Respironics, Inc.
REF 580-1000-02 G
Oxygen Supply
(Test 4, Step 2)
This step prompts the operator to disconnect oxygen to verify that the oxygen
pressure switch (PS1) opens.
Diagnostic code
3112
indicates that PS1 is closed. This can occur if oxygen is
not connected, or if the PS1 connection or the sensor PCB is faulty.
Red = air flow
Blue = O
2
flow
Black = no flow
Содержание ESPRIT
Страница 1: ...REF 580 1000 02 G ESPRIT Ventilator S e r v i c e M a n u a l ...
Страница 318: ...Appendix A A 24 Esprit Ventilator Service Manual Respironics Inc REF 580 1000 02 G This page is intentionally blank ...
Страница 320: ...Appendix B B 2 Esprit Ventilator Service Manual Respironics Inc REF 580 1000 02 G This page is intentionally blank ...