Appendix A
REF 580-1000-02 G
Esprit® Ventilator Service Manual © Respironics, Inc.
A-5
Safety Valve
(Test 2, Step 1)
This step charges the circuit to verify that the system is closed.
Diagnostic code
3106
occurs if the ventilator cannot pressurize the system to
50 cmH
2
O.
Red = air flow
Blue = O
2
flow
Black = no flow