Chapter 9
Component Removal/Installation
REF 580-1000-02 G
Esprit® Ventilator Service Manual © Respironics, Inc.
9-81
Figure 9-89: Cooling Coil and Fan
GUI Upgrades
Three GUI versions exist in the field: the original monochrome 9.5-in. GUI, the
original 10.4-in. GUI, and the 2nd generation 10.4-in. GUI. Because the 9.5-
in. GUI assembly is no longer available, the 10.4-in. GUI upgrade kit must be
installed when installing the color upgrade option or replacing a VGA display,
touch frame, backlight inverter PCB, or keyboard.
The GUI upgrade procedure includes the following steps, if applicable:
•
Replacing the GUI (see page 9-84).
•
Installing the upgrade kit CPU and DC/DC converter PCBs (see page 9-
87).
•
Downloading ventilator software (see page 9-87).
•
Calibrating the screen (see page 9-88).
•
Enabling the color option (see page 9-88).
•
Enabling the graphics option (see page 9-88).
•
Final checkout (see page 9-88).
Cooling coil (note orientation of inlet)
Fan gasket
positioned inside
bottom enclosure
Four screws, washers
(view from outside
bottom enclosure)