Chapter 9
Component Removal/Installation
REF 580-1000-02 G
Esprit® Ventilator Service Manual © Respironics, Inc.
9-35
Figure 9-37: LCD (2nd Generation)
GUI Cleaning and
Dust Gasket
Installation (2nd
Generation 10.4-in.
GUI)
Follow these steps to clean the LCD and IR TouchFrame.
1. Remove the GUI (see “Backlight Inverter PCB, VGA Display, and
Touch Frame (Original 10.4-in. GUI)” on page 9-32, steps 1-6).
2. To remove the LCD and transition PCBA, remove the two M3 x 16-mm
screws and two M3 flat washers that hold the LCD to the LCD bracket
(Figure 9-36).
3. Set the LCD face up on a nonabrasive surface.
4. Clean the IR TouchFrame and LCD assemblies with isopropyl alcohol
with a soft, lint free cloth.
5. Separate the GUI dust gasket parts (Figure 9-38).
6. Place the four foam pieces on the LCD (Figure 9-39 and Figure 9-40).
7. Verify the two 5/16 hex spacers are aligned on the LCD bracket.
8. Reinstall the LCD and transition PCBA onto the bezel. Use four M3 x
5-mm screws and four M3 flat washers to attache the LCD. Then use
Содержание ESPRIT