Chapter 9
9-20
Esprit® Ventilator Service Manual © Respironics, Inc.
REF 580-1000-02 G
Component Removal/Installation
3. Remove the screws securing the power supply to the top enclosure.
4. Move the power supply out of the way to allow access to the GUI
mounting screws.
5. Remove the sensor PCB (see “Sensor PCB” on page 9-5).
6. Loosen the set screws (one for each knob) holding the brightness and
volume knobs.
7. Remove four 3-mm screws securing the GUI to the top enclosure.
8. Carefully remove the GUI assembly from the top enclosure.
Figure 9-22: 9.5-in. GUI Assembly
Follow these steps to install the 9.5-in. GUI assembly (Figure 9-23):
1. Assemble the knobs to brightness and volume potentiometers so that
set screws are closest to potentiometers. On the intensity
potentiometer, ensure that the hex screw on the knob makes contact
with the flat part of the shaft.
2. Feed cables 1 through 5 through the top enclosure slots, then set the
GUI assembly in position onto the top enclosure, ensuring that all
ribbon cables are untwisted.
3-mm screws