Chapter 9
9-18
Esprit® Ventilator Service Manual © Respironics, Inc.
REF 580-1000-02 G
Component Removal/Installation
MMI PCB
(10.4-in. GUI)
Follow these steps to remove the MMI PCB (Figure 9-21) for ventilators with
10.4-in. GUI. Reverse removal steps to install.
1. Remove the top enclosure (see “Top Enclosure” on page 9-4).
2. Remove the power supply shroud (see “Power Supply Fan/Shroud” on
3. Remove the screws securing the power supply to the top enclosure.
4. Move the power supply out of the way to allow access to the MMI
mounting screws.
5. Remove the sensor PCB (see “Sensor PCB” on page 9-5).
6. Remove the cables from the connectors on the MMI PCB:
Remove cables in this order:
From this connector on
the MMI PCB (10.4-in.
GUI):
MMI J1
POWER SWITCH
J1
MMI J3
MAIN HARNESS
J3
MMI J4
PRIMARY ALARM
J4
MMI J5
ALARM VOLUME POT
J5
Ribbon cable from keyboard LEDs*
J6
MMI J8
MAIN PCB
J8
Rotary encoder ribbon cable
J10
Ribbon cable from keyboard overlay LEDs*
(Newer J11 connectors are a friction fit,
therefore a connector lock is not present.)
J11
White ribbon cable from touch frame*
Display bottom
J12
White ribbon cable from touch frame*
Display top
J13
MMI J15
BACKUP ALARM
J15
* When removing, lift connector lock before pulling cable. Do not use a tool.