Chapter 9
Component Removal/Installation
REF 580-1000-02 G
Esprit® Ventilator Service Manual © Respironics, Inc.
9-13
5. Fasten the black ground wire with longer screw (M5x10) to the shroud
near the TB2 terminal block.
Figure 9-16: Shroud Ground Connection
6. Fasten the shroud with the remaining four M5x6 screws.
7. Connect the cables to the backlight inverter PCB (for the 9.5-in. GUI)
or DC/DC converter PCB (10.4-in. GUI).
8. Reinstall the top enclosure (see “Top Enclosure” on page 9-4).
Power Supply Fuses
WARNING:
The power supply generates high voltage. To avoid personal injury, verify that
the AC and external DC power sources are disconnected from the ventilator.
NOTE:
F2 fuses will
not
be field replaceable on the power supplies with fuse F2
soldered to the fuse clip (see Figure 9-17). F2 fuses
will
be available for
replacement in Esprit ventilators with power supplies without fuse F2
soldered to the fuse clip (see Figure 9-18).
1. Remove the top enclosure (see “Top Enclosure” on page 9-4).
2. Remove the power supply shroud (“Power Supply Fan/Shroud” on
3. Remove the appropriate fuse from its holder (see Figure 9-17).
4. Pinch fuse clips together before inserting replacement fuse (see
TB2
Shroud
M5x10
screw