Chapter 9
9-12
Esprit® Ventilator Service Manual © Respironics, Inc.
REF 580-1000-02 G
Component Removal/Installation
Figure 9-15: Power Supply Connectors
NOTE:
The GUI assembly and MMI PCB must be installed before reinstalling the
power supply on ventilators with the 9.5-in. GUI. The MMI PCB must be
installed on ventilators with the 10.4-in. GUI before reinstalling the
power supply. When reinstalling, ensure that no cables are routed
underneath the power supply.
Follow these steps to reinstall the power supply:
1. Reinstall the eight screws that hold the power supply to the top
enclosure. Do not tighten the screws until all are installed.
2. Connect these cables to the power supply:
•
Cable PSU J1 to connector J1.
•
Cable PSU J2 to connector J2.
•
Cable PSU J3 to connector J3.
3. Install these wires to the power supply, pushing the wires completely
into the connectors. No wires should be visible after being tightened at
the terminal block.
•
Red and black wires (PSU TB3) to connector TB3.
•
Red and black wires (PSU TB2) to connector TB2.
•
Blue, green, and brown wires from the AC mains panel to
connector TB1.
•
Connect the fan harness.
4. Replace the power supply shroud, ensuring that the red and black
wires from TB3 rest inside the shroud slots, and that the shroud does
not pinch any of these cables. All cable assemblies and wires must be
routed between the shroud and the top enclosure wall.