Chapter 8
Performance Verification
REF 580-1000-02 G
Esprit® Ventilator Service Manual © Respironics, Inc.
8-43
Test 3: Air Flow Accuracy
Test 4: Oxygen Flow Accuracy
Symptom
Recommended Repair
Measured flow rates out
of range
1. Confirm correct altitude setting.
2. Check that safety valve solenoid is energized.
3. Verify calibration analyzer is set to measure air in ATP mode.
4. Check that nothing is obstructing the ventilator inspiratory outlet.
5. Verify that CV3 is correctly positioned and oriented.
6. If air flow sensor reading is within specification of the calibration
analyzer, check for exhalation bacteria filter occlusions (perform
the backpressure test or replace filter) and confirm CV4
orientation before replacing the exhalation flow sensor.
7. If air flow sensor reading is not within limits as compared to the
calibration analyzer, replace the air flow sensor.
8. Slave in a different air flow sensor cable.
9. Replace air motor controller PCB.
10.Replace the sensor PCB.
11.Replace the air valve assembly.
Symptom
Recommended Repair
Measured flow rates out
of tolerance
1. Confirm correct altitude setting.
2. Check that safety valve solenoid is energized.
3. Verify calibration analyzer is set to measure oxygen in ATP mode.
4. Check that nothing is obstructing the ventilator inspiratory outlet.
5. Verify that CV3 is correctly positioned and oriented.
6. If oxygen flow sensor reading is within specification of the
calibration analyzer, check for exhalation bacteria filter occlusions
(perform the backpressure test or replace filter) and confirm CV4
orientation before replacing the exhalation flow sensor.
7. If oxygen flow sensor reading is not within limits as compared to
the calibration analyzer, replace the oxygen flow sensor.
8. Slave in a different oxygen flow sensor cable.
9. Replace oxygen motor controller PCB.
10.Replace the sensor PCB.
11.Replace the oxygen valve assembly.