
Chapter 8
Performance Verification
REF 580-1000-02 G
Esprit® Ventilator Service Manual © Respironics, Inc.
8-21
7. Touch
Exhalation
and adjust the steps until the analyzer pressure
reads 100 ± 5 cmH
2
O (95 to 105 cmH
2
O).
8. Check that the Inhalation Pressure and Exhalation Pressure displays
on the hardware screen read within ± 10% of the analyzer display.
EXAMPLE: If the analyzer’s display reads 98.7 cmH
2
O, 10% of the
display is 9.87 cmH
2
O, and ± 10% of the analyzer display would be
88.83 to 108.57 cmH
2
O (-10% = 98.7 - 9.87 = 88.83 cmH
2
O, and
+ 10% = 98.7 + 9.87 = 108.57 cmH
2
O).
9. The pressure accuracy test is complete.
Figure 8-20: Pressure Accuracy Test Configuration
PEEP System (Test 6)
The PEEP system test verifies the integrity of the PEEP system and checks for
autocycling.
Содержание ESPRIT
Страница 1: ...REF 580 1000 02 G ESPRIT Ventilator S e r v i c e M a n u a l ...
Страница 318: ...Appendix A A 24 Esprit Ventilator Service Manual Respironics Inc REF 580 1000 02 G This page is intentionally blank ...
Страница 320: ...Appendix B B 2 Esprit Ventilator Service Manual Respironics Inc REF 580 1000 02 G This page is intentionally blank ...