Doc 0166440 rev. E
Page 11 of 19
Test Equipment - MATCH 70/80
CS44 Prog unit
p/n 16687500
CS44 prog cable
Test fixture, p/n 16936600
Attach the CS44 prog
cable to the prog unit
Set trimmers to max. Re-
move batt door and insert
prog unit into the HI. Align the
interior hinge of the batt door
from the HI with the hinge
groove on the prog unit
Hold prog unit down
onto the HI by pressing
the top portion of the
prog unit as shown
While holding the unit
down onto the HI,
slide the unit with HI
onto the test fixture
Slide the prog cable on the slot
of the fixture as shown. Tighten
the bolt screw to keep the HI and
prog unit in place (turn clockwise)
Attach the HI to the 2cc
coupler and place it in the
test chamber as shown
Pressure
points
Pressure
point
Bolt screw to
tighten